Agilent InfinityLab LC Series Vialsamplers User Manual 348
12 LAN Configuration
PC and Agilent ChemStation Setup
Agilent ChemStation Setup
This procedure describes the Agilent ChemStation B.04.02 setup for the 1290
Infinity system using the 1290 Infinity DAD (G4212A) as the interfacing module.
The setup works in the same way for all other systems.
1 Open the ChemStation Configuration Editor.
2 Select from the menu Configure - Instruments.
3 Select Modular 3D LC System.
4 Give the Instrument a name.
NOTE
The LAN must be connected to detector due to high data load on communication
to Control Software.