Agilent InfinityLab LC Series Vialsamplers User Manual 135
3Using the Module
Transporting the Sampler
Install the Transport Protection Foam
When
To secure the transport arm before transporting or shipping the sampler.
Parts required p/n Description
G7129-40050 Transport Protection Foam
Preparations
All vials are removed from the vial drawers.
1 Open the doors of the module. 2 Move the gripper arm into the park position (see “Park
Arm on page 186) and turn off the sampler.
3 Place the Transport Protection Foam into the sampler. 4 Push the Transport Foam in until the endpoint.