Agilent InfinityLab LC Series Vialsamplers User Manual 34
2 Site Requirements and Specifications  
Site Requirements
Site Requirements
Site Requirements
A suitable environment is important to ensure optimum performance of the 
instrument.
Power Considerations
The module power supply has wide ranging capability. It accepts any line voltage 
in the range described in 
Table 1 on page 38. Consequently there is no voltage 
selector in the rear of the module. There are also no externally accessible fuses, 
because automatic electronic fuses are implemented in the power supply.
Hazard of electrical shock or damage of your instrumentation
can result, if the devices are connected to a line voltage higher than 
specified.
 Connect your instrument to the specified line voltage only.
Electrical shock hazard
The module is partially energized when switched off, as long as the power 
cord is plugged in.
The cover protects users from personal injuries, for example electrical 
shock.
 Do not open the cover.
 Do not operate the instrument and disconnect the power cable in case 
the cover has any signs of damage.
 Contact Agilent for support and request an instrument repair service.
Inaccessible power plug.
In case of emergency it must be possible to disconnect the instrument 
from the power line at any time.
 Make sure the power connector of the instrument can be easily reached 
and unplugged.
 Provide sufficient space behind the power socket of the instrument to 
unplug the cable.