Agilent InfinityLab LC Series Vialsamplers User Manual 200
7Error Information
General Error Messages
Fan Failed
Error ID: 0068
The fan in the autosampler module or in the Sample Cooler/Sample Thermostat
has failed.
The hall sensor on the fan shaft is used by the mainboard to monitor the fan
speed. If the fan speed falls below a certain limit for a certain length of time, the
error message is generated.
This limit is given by 2 revolutions/second for longer than 5 seconds.
Depending on the module, assemblies (e.g. the lamp in the detector) are turned
off to assure that the module does not overheat inside.
Probable cause Suggested actions
1
Fan cable disconnected.
Please contact your Agilent service representa-
tive.
2
Defective fan.
Please contact your Agilent service representa-
tive.
3
Defective Sample Cooler/Sample Thermo-
stat fan
Replace the Sample Cooler/Sample Thermostat.
4
Defective mainboard.
Please contact your Agilent service representa-
tive.