Agilent InfinityLab LC Series Vialsamplers User Manual 39
2 Site Requirements and Specifications  
Performance Specifications
Performance Specifications
Performance Specifications (G7129A)
Table 2 Performance Specifications G7129A
Type Specification Comments Method/Conditions
Injection range 0.1 – 100 µL in 0.1 µL incre-
ments with 100 µL up to 
60
 MPa
0.1 – 900 µL in 0.1 µL incre-
ments with 900 µL up to 
40
 MPa
Up to 1800 µL with multiple 
draw (hardware modification 
required)
Requires 900 µL analytical head
Injection precision <0.25 % RSD of peak areas from 
5
 µL to 100 µL
Measured caffeine
Pressure range 0 – 60 MPa (0 – 600 bar, 
0
 – 8702 psi)
0 – 40 MPa (0 – 400 bar, 
0
 – 5801 psi)
for 900 µL Analytical Head
Sample viscosity 
range
0.2 – 5.0 cP
Sample capacity 132 x 2 mL vial (two trays 
default)
100 x 2 mL vial (two classic 
trays optional)
36 x 6 mL vials (two trays 
optional)