Agilent InfinityLab LC Series Vialsamplers User Manual 186
6 Troubleshooting and Diagnostics
Maintenance and Troubleshooting Tools
Park Arm
For transportation of the Vialsampler, the transport arm must be moved into the
park position and secured with the transport foam ( Transport Protection Foam
(G7129-40050)).
The Park Arm function can be found under Special Commands in the Instrument
Control tab of the Agilent Lab Advisor software.
CAUTION
Unsecured transportation of the module
The Vialsampler must be transported in the park position with the transport
foam installed. Failing to do so might lead to damage to the internal parts.
Ensure that all vials are removed from the vial drawers.
Ensure that the transport assembly is in the park position.
Ensure that the Transport Protection Foam (G7129-40050) is correctly
installed.