Agilent InfinityLab LC Series Vialsamplers User Manual 52
2 Site Requirements and Specifications
Specifications of the Integrated Column Compartment
Performance Specifications Agilent InfinityLab LC
Series Integrated Column Compartment (G7130A)
Table 10 Performance Specifications G7130A
Type Specification Comments
Temperature range 5 °C above ambient to 80 °C
Column capacity 2 columns up to 30 cm and
4.6
mm ID
Temperature stability ±0.10 °C at sensor
Temperature accuracy ±0.8 K (±0.5 K with calibration)
Warm up time 2040 °C in 5 min at sensor