Agilent InfinityLab LC Series Vialsamplers User Manual 208
7Error Information  
Vialsampler Error Messages
Cleanup after abort failed
Error ID: 34310
This, similarly to the Command failed error events, is a non-specific error message 
and indicates only that something went wrong during the cleanup procedure. The 
cleanup procedure refers to the process, which is executed when an analysis is 
aborted (for example, due to a missing vial) and the components of the sampler 
(such as the needle, gripper arm, and the metering device) need to be reset to the 
initial conditions. As for the Command failed error events, this error message is 
usually also coupled with other error events, which provide more specific 
information on the nature of the issue. To identify these source error events, look 
for entries with the same timestamp as for the Cleanup after abort failed error 
message.
Wash procedure failed
Error ID: 34314
This, similarly to the Command failed error events, is a non-specific error message 
and indicates only that something went wrong during the wash procedure. It is 
usually coupled with other error events, which provide more specific information 
on the nature of the issue. To identify these source error events, look for entries 
with the same timestamp as for the Wash procedure failed error message.
Probable cause Suggested actions
1
Cleanup procedure failed
Identify the source error message and follow the 
instructions provided for that.
Probable cause Suggested actions
1
Wash procedure failed
Identify the source error message and follow the 
instructions provided for that.