2 Site Requirements and Specifications
Specifications of the Sample Cooler
Specifications of the Sample Cooler
The Agilent Infinity II Sample Cooler is a vapor-compression refrigeration system
that uses a fluorinated greenhouse gas (HCF-134a) as the refrigerant. For
information on carbon dioxide equivalency (CDE) and global warming potential
(GWP), see the instrument label.
Table 5 Physical Specification of the Sample Cooler
Type Specification Comment
Weight < 6 kg (< 13.2 lbs)
Dimensions (height × width × depth) 205 x 340 x 370 mm
(8.1 x 13.4 x 14.6 inches)
Refrigerant gas HFC-134a (0.042 kg) Ozone depletion poten-
tial (ODP) = 0
Supply voltage 24 VDC
Current 10 A max.
Ambient operating temperature 4 – 40 ° C (39104 ° F)
Ambient non-operating temperature -4070 ° C (-40158 ° F)
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Safety standards:
IEC, EN, CSA, UL
Installation category II, Pollution
degree 2
For indoor use only
ISM Classification ISM Group 1 Class B According to CISPR 11