Agilent InfinityLab LC Series Vialsamplers User Manual 142
4 Preparing the Module
Solvent Information
Solvent Information
Observe the following recommendations on the use of solvents.
Follow the recommendations for avoiding the growth of algae, see the pump
manuals.
Small particles can permanently block capillaries and valves. Therefore,
always filter solvents through 0.22
µm filters.
Avoid or minimize the use of solvents that may corrode parts in the flow path.
Consider specifications for the pH range given for different materials such as
flow cells, valve materials etc. and recommendations in subsequent sections.
Recommended Wash Solvents
water
ethanol
methanol
water/acid (especially for basic compounds)
water/base (especially for acidic compounds)
water/acetonitrile
NOTE
For different wash solvents as mentioned above, verify that the wash solvent is
suitable for the silicone wash tubing.

Table of Contents

Other manuals for Agilent Technologies InfinityLab LC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfinityLab LC Series and is the answer not in the manual?

Agilent Technologies InfinityLab LC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinityLab LC Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals