EasyManuals Logo

Agilent Technologies InfinityLab LC Series User Manual

Agilent Technologies InfinityLab LC Series
369 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #197 background imageLoading...
Page #197 background image
Agilent InfinityLab LC Series Vialsamplers User Manual 197
7Error Information
General Error Messages
Leak Sensor Open
Error ID: 0083
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current falls outside the lower limit, the error
message is generated.
Probable cause Suggested actions
1
Leak sensor not connected to the power
switch board.
Please contact your Agilent service representa-
tive.
2
Defective leak sensor.
Please contact your Agilent service representa-
tive.
3
Leak sensor incorrectly routed, being
pinched by a metal component.
Please contact your Agilent service representa-
tive.
4
Power switch assembly defective
Please contact your Agilent service representa-
tive.

Table of Contents

Other manuals for Agilent Technologies InfinityLab LC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfinityLab LC Series and is the answer not in the manual?

Agilent Technologies InfinityLab LC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinityLab LC Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals