Agilent InfinityLab LC Series Vialsamplers User Manual 197
7Error Information
General Error Messages
Leak Sensor Open
Error ID: 0083
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current falls outside the lower limit, the error
message is generated.
Probable cause Suggested actions
1
Leak sensor not connected to the power
switch board.
Please contact your Agilent service representa-
tive.
2
Defective leak sensor.
Please contact your Agilent service representa-
tive.
3
Leak sensor incorrectly routed, being
pinched by a metal component.
Please contact your Agilent service representa-
tive.
4
Power switch assembly defective
Please contact your Agilent service representa-
tive.