2 Site Requirements and Specifications  
Performance Specifications
Performance Specifications (G7129B)
Table 3 Performance Specifications G7129B
Type Specification Comments Method/Conditions
Injection range 0.1 – 20 µL in 0.1 µL increments 
(default)
0.1 – 40 µL in 0.1 µL increments 
(if 40
 µL loop is installed)
0.1 – 120 µL in 0.1 µL incre-
ments with 1290 Infinity large 
volume injection kit (hardware 
modification required)
0.1 – 100 µL in 0.1 µL (if 100 µL 
loop and 100
 µL head is 
installed)
Up to 1500 µL (with 1400 µL 
multi-draw kit and 100
 µL ana-
lytical head)
up to 130 MPa (1300 bar, 
18854
 psi)
up to 60 MPa (600 bar, 
8702
 psi)
Injection precision <0.25 % RSD of peak areas from 
5
 µL to 100 µL
Measured caffeine
Pressure range Up to 130 MPa (1300 bar, 
18854
 psi)
Sample viscosity 
range
0.2 – 5.0 cP
Sample capacity 132 x 2 mL vial (two trays 
default)
100 x 2 mL vial (two classic 
trays optional)
36 x 6 mL vials (two trays 
optional)