2 Site Requirements and Specifications
Physical Specifications
Physical Specifications
Table 1 Physical Specifications G7129A, G7129B, G7129C
Type Specification Comments
Weight 19 kg (41.9 lbs) w/o sample thermostat
Dimensions
(height × width × depth)
320 x 396 x 468 mm
(12.8 x 15.6 x 18.4 inches)
Line voltage 100240 V~, ± 10 % Wide-ranging capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 350 VA, 350 W, 1195 BTU/h
Ambient operating tempera-
ture
4 - 40 °C (39 - 104 °F); without sample ther-
mostat up to 55 °C (131 °F)
Ambient non-operating tem-
perature
-4070 °C (-40158 °F)
Humidity
< 95 % r.h. at 40 °C (104 °F)
1
1
If a sample thermostat is included the upper value for humidity can be reduced. Please check your lab
conditions to stay beyond dew point values for non–condensing operation.
Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Safety standards:
IEC, EN, CSA, UL
Installation category II, Pollution degree 2 For indoor use only.
ISM Classification ISM Group 1 Class B According to CISPR 11
Permitted solvents Boiling point 56 °C (133 °F).
Auto-ignition temperature 200 °C
(394
°F).

Table of Contents

Other manuals for Agilent Technologies InfinityLab LC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfinityLab LC Series and is the answer not in the manual?

Agilent Technologies InfinityLab LC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinityLab LC Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals