Agilent InfinityLab LC Series Vialsamplers User Manual 38
2 Site Requirements and Specifications
Physical Specifications
Physical Specifications
Table 1 Physical Specifications G7129A, G7129B, G7129C
Type Specification Comments
Weight 19 kg (41.9 lbs) w/o sample thermostat
Dimensions
(height × width × depth)
320 x 396 x 468 mm
(12.8 x 15.6 x 18.4 inches)
Line voltage 100 – 240 V~, ± 10 % Wide-ranging capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 350 VA, 350 W, 1195 BTU/h
Ambient operating tempera-
ture
4 - 40 °C (39 - 104 °F); without sample ther-
mostat up to 55 °C (131 °F)
Ambient non-operating tem-
perature
-40 – 70 °C (-40 – 158 °F)
Humidity
< 95 % r.h. at 40 °C (104 °F)
1
1
If a sample thermostat is included the upper value for humidity can be reduced. Please check your lab
conditions to stay beyond dew point values for non–condensing operation.
Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Safety standards:
IEC, EN, CSA, UL
Installation category II, Pollution degree 2 For indoor use only.
ISM Classification ISM Group 1 Class B According to CISPR 11
Permitted solvents Boiling point ≥56 °C (133 °F).
Auto-ignition temperature ≥200 °C
(394
°F).