Agilent InfinityLab LC Series Vialsamplers User Manual 286
9 Parts and Materials for Maintenance  
Multidraw Kits
Multidraw Kits
Multidraw Kit
Multidraw Kit for InfinityLab Sampler (G7167-68711) contains the following 
items:
Large Volume Injection Kit
Large Volume Injection Kit for 1290 Infinity II Samplers (G4216-68711) contains:
p/n Description
G1313-87307 Seat Capillary, ST, 500 µL, ID 0.5 mm
G1313-87308 Seat Capillary, ST, 1500 µL, ID 0.94 mm
0100-0900 Zero Dead Volume Union, ST
With the Multidraw Kit, the maximum injection volume can be extended to 500 or 
1500 µL for the 100 µL analytical head and to 1800 µL for the 900 µL analytical 
head. The nominal volume of the seat capillaries is 656 and 2415 µL, 
respectively.
p/n Description
Extension Seat Capillary, 80 µL, 0.5 mm ID (0.9 mm OD)
With the Large Volume Injection Kit, the maximum injection volume can be 
extended to 100 or 120 µL, depending on the sample loop installed in the system. 
The nominal volume of the seat capillary is 157 µL.