EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>InfinityLab LC Series

Agilent Technologies InfinityLab LC Series User Manual

Agilent Technologies InfinityLab LC Series
369 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #173 background imageLoading...
Page #173 background image
Agilent InfinityLab LC Series Vialsamplers User Manual 173
5 Optimizing Performance
Precise Injection Volume
Precise Injection Volume
Injection Volumes Less Than 2 µL
When the injection valve switches to the BYPASS position, the mobile phase in
the sample loop is depressurized. When the syringe begins drawing sample, the
pressure of the mobile phase is decreased further. If the mobile phase is not
degassed adequately, small gas bubbles can form in the sample loop during the
injection sequence. When using injection volumes <
2 µL, these gas bubbles can
affect the injection-volume precision. Use degassed mobile phases for best
injection-volume precision with injection volumes <
2 µL.
If the pump has no degasser, it is recommended to use an Agilent 1260 Infinity
degasser.
Also, using the automated needle wash (see “Optimization for Lowest
Carryover” on page 167) between injections reduces carryover to a minimum,
further improving the injection volume precision.

Table of Contents

Other manuals for Agilent Technologies InfinityLab LC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfinityLab LC Series and is the answer not in the manual?

Agilent Technologies InfinityLab LC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinityLab LC Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals