EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>InfinityLab LC Series

Agilent Technologies InfinityLab LC Series User Manual

Agilent Technologies InfinityLab LC Series
369 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #212 background imageLoading...
Page #212 background image
Agilent InfinityLab LC Series Vialsamplers User Manual 212
7Error Information
Vialsampler Error Messages
Timeout during axis movement
Error ID: 34330
The system reported an error because the execution of the given transport
command took longer than expected.
34330,0 corresponds to the X-axis movement.
34330,1 corresponds to the Z-axis movement.
34330,2 corresponds to the theta movement.
34330,3 corresponds to the gripper finger movements.
Probable cause Suggested actions
1
Contaminated transport rods
Inspect the x- and z-axis transport rods and clean
them if you observe contamination or dust accu
-
mulation on them. Use isopropanol and lint-free
cloth for this purpose.
2
Broken gripper
Verify the correct functioning of the gripper by
using the Gripper Verification tool. Replace the
gripper if needed.
3
Potential hardware error
Please contact your Agilent service representa-
tive.
4
Misaligned gripper arm
Verify that the alignment parameters are in the
correct range and perform the Automatic Refer
-
encing if needed.

Table of Contents

Other manuals for Agilent Technologies InfinityLab LC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfinityLab LC Series and is the answer not in the manual?

Agilent Technologies InfinityLab LC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinityLab LC Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals