EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>InfinityLab LC Series

Agilent Technologies InfinityLab LC Series User Manual

Agilent Technologies InfinityLab LC Series
369 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #172 background imageLoading...
Page #172 background image
Agilent InfinityLab LC Series Vialsamplers User Manual 172
5 Optimizing Performance
Fast Injection Cycle and Low Delay Volume
General Recommendations for Fast Injection Cycle
Times
As described in this section, the first step to provide short cycle times are
optimizing the chromatographic conditions. If this is done the autosampler
parameter should be set to:
Overlapped injection mode
Increase of draw and eject speed for large injection volumes
Add at last run a blank, if overlapped injection is used
To reduce the injection time, the detector balance has to be set to OFF.