5 Optimizing Performance  
Precise Injection Volume
Draw and Eject Speed
Draw Speed
The speed at which the metering unit draws sample out of the vial may have an 
influence on the injection volume precision when using viscous samples. If the 
draw speed is too high, air bubbles may form in the sample plug, affecting 
precision. The default draw speed is 100
 µL/min. This speed is suitable for the 
majority of applications, however, when using viscous samples, set the draw 
speed to lower speed for optimum results. A DRAW statement in an injector 
program also uses the draw speed setting which is configured for the 
autosampler.
Eject Speed
The default eject speed setting is 400 µL/min. When using large injection 
volumes, setting the eject speed to a higher value speeds up the injection cycle 
by shortening the time the metering unit requires to eject solvent at the beginning 
of the injection cycle (when the piston returns to the home position).
An EJECT statement in an injector program also uses the eject speed setting 
which is configured for the autosampler. A faster eject speed shortens the time 
required to run the injector program. When using viscous samples, a high eject 
speed should be avoided.