Agilent InfinityLab LC Series Vialsamplers User Manual 317
11 Hardware Information  
Electrical Connections
Electrical Connections
• The CAN bus is a serial bus with high-speed data transfer. The two 
connectors for the CAN bus are used for internal module data transfer and 
synchronization.
• The ERI/REMOTE connector may be used in combination with other analytical 
instruments from Agilent Technologies if you want to use features such as 
start, stop, common shutdown, prepare, and so on.
• With the appropriate software, the LAN connector may be used to control the 
module from a computer through a LAN connection. This connector is 
activated and can be configured with the configuration switch.
• With the appropriate software, the USB connector may be used to control the 
module from a computer through a USB connection.
• The power input socket accepts a line voltage of 100 – 240 VAC ± 10 % with a 
line frequency of 50 or 60
 Hz. Maximum power consumption varies by 
module. There is no voltage selector on your module because the power 
supply has wide-ranging capability. There are no externally accessible fuses 
because automatic electronic fuses are implemented in the power supply.
Never use cables other than the ones supplied by Agilent Technologies to ensure 
proper functionality and compliance with safety or EMC regulations.