Agilent InfinityLab LC Series Vialsamplers User Manual 88
3Using the Module
Install the Optional Integrated Column Compartment
Install the Column Shelf
Tools required p/n Description
8710-0510 Open-end wrench 1/4 — 5/16 inch
Parts required p/n Description
G7129-60023 Column Shelf
Preparations
Switch off the pump.
Switch off the sampler.
Open the doors of the sampler.
Open the ICC flap door.
WARNING
Fire or explosion due to missing leak drainage
Solvents may run into the module if column shelf or Integrated Column
Compartment are not installed.
Install either column shelf or Integrated Column Compartment before
starting the module.
1 Remove the capillary connecting the injection valve and
the ICC heat exchanger.
2 Remove the column from the column holder clips.