EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>InfinityLab LC Series

Agilent Technologies InfinityLab LC Series User Manual

Agilent Technologies InfinityLab LC Series
369 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #168 background imageLoading...
Page #168 background image
Agilent InfinityLab LC Series Vialsamplers User Manual 168
5 Optimizing Performance
Optimization for Lowest Carryover
Using the Needle Wash in the Wash Vial
For very critical applications where the outside of the needle cannot be cleaned
sufficiently with one wash port solvent you can use an injector program and an
additional wash vial with an appropriate and stronger solvent for cleaning. The
wash vial should contain solvent in which the sample components are soluble,
and the vial should not be capped. If the wash vial is capped, small amounts of
sample remain on the surface of the septum, which may be carried on the needle
to the next sample.
Injector Program with Needle Wash
The injector program includes the command WASH. When this command is
included in the injector program, the needle is lowered once into the specified
wash vial or flushport before injection.
For example:
1 WASH > Location P2-A1 or flushport
2 Draw 5 µL from Sample Location P1-A1
3 INJECT
Line 1 moves the needle to vial P2-A1 or the flushport. Line 2 draws 5 µL from the
current sample vial P1-A1. Line 3 injects the sample (valve switches to main
pass).

Table of Contents

Other manuals for Agilent Technologies InfinityLab LC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfinityLab LC Series and is the answer not in the manual?

Agilent Technologies InfinityLab LC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinityLab LC Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals