Agilent InfinityLab LC Series Vialsamplers User Manual 230
8Maintenance
Exchange the Needle Assembly
WARNING
Risk of injury by uncovered needle
An uncovered needle is a risk of harm to the operator.
Do not open the safety cover of the needle station during normal
operation.
Wear safety goggles and safety gloves when removing the needle
assembly.
NOTE
It is recommended to always exchange the needle assembly and the needle seat
at the same time to prevent premature leakage.
NOTE
When the instrument setup has changed, configure the new setup in the online
view of the Chromatographic Data System. See Table 15 on page 155.
1 In the Agilent Lab Advisor software select Service &
Diagnostics> Maintenance Positions> Change Needle,
Loop and Seat, click Start and wait until the needle
assembly is in maintenance position.
OR
In the Local Controller start the maintenance mode and
select Change Needle, Loop and Seat function.
2 Open the doors of the module.

Table of Contents

Other manuals for Agilent Technologies InfinityLab LC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfinityLab LC Series and is the answer not in the manual?

Agilent Technologies InfinityLab LC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinityLab LC Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals