Agilent InfinityLab LC Series Vialsamplers User Manual 232
8Maintenance
Exchange the Needle Assembly
7 Disconnect the sample loop from the needle, and, if
needed, also from the metering device.
8 Loosen the fixing screw.
9 Lift out the needle. 10 Mount and align the needle on its fixture, then tighten
the screw firmly.