Agilent InfinityLab LC Series Vialsamplers User Manual 242
8Maintenance
Exchange the Sample Loop Assembly
1 In the Agilent Lab Advisor software select Service &
Diagnostics> Maintenance Positions> Change Needle,
Loop and Seat, click Start and wait until the needle
assembly is in maintenance position.
OR
In the Local Controller start the maintenance mode and
select Change Needle, Loop and Seat function.
2 Open the doors of the module.
3 Loosen the safety cover screw.
WARNING
Sharp needle
Uncovered needles may cause injuries
Do not touch the tip of the needle.
4 Remove the safety cover.