Agilent InfinityLab LC Series Vialsamplers User Manual 257
8Maintenance
Replace the Analytical Heads
10 Close the doors.
Next Steps:
11 Perform a System Pressure Test.
12 If needed, install a sample loop matching the volume of
the new analytical head (see
“Exchange the Sample
Loop Assembly” on page 241).