Agilent InfinityLab LC Series Vialsamplers User Manual 308
10 Identifying Cables
Remote Cables
5188-8057 ERI to APG and RJ45 (Connector D_Subminiature 15 pin (ERI),
Connector D_Subminiature 9 pin (APG), Connector plug Cat5e (RJ45))
One end of these cables provides a Agilent Technologies APG (Analytical
Products Group) remote connector to be connected to Agilent modules. The
other end depends on the instrument to be connected to.
Table 22 5188-8057 ERI to APG and RJ45
p/n 5188-8057 Pin (ERI) Signal Pin (APG) Active (TTL) Pin (RJ45)
10 GND 1 5
1 Start
Request
9 High
2 Stop 8 High
3 Ready 7 High
4 Fraction Trig-
ger
5 High 4
5 Power on 6 High
6 Shut Down 4 High
7 Start 3 High
8 Prepare 2 High
Ground Cable Shield-
ing
NC