Chapter 3 143
Mode
Adjacent Channel Power—ACP (I&M) Measurement
Mode
Adj Chan Meas BW Sets the adjacent channel measurement bandwidth.
Adj Chan Meas Time Sets the adjacent channel measurement time. This feature is mainly for
bursted signals. The sweep time has to be defined such that it does cover a complete frame or period
of the signal.
Freq Error Limits On/Off Turns the frequency error limits for the adjacent channel power
measurement On or Off.
Mode Channel Analyzer
Dependencies /
Couplings
Format Type must be set to Cust
Preset 30 kHz
State Saved Saved in instrument state.
Min 10 kHz
Max 5 MHz
Key Path
Meas Setup, Format/BW, Channel BW/Time/Freq Err, Adj Chan Meas
BW
Mode Channel Analyzer
Dependencies /
Couplings
Format Type must be set to Cust
Preset 10 ms
State Saved Saved in instrument state.
Min BW dependent
Max BW dependent
Key Path
Meas Setup, Format/BW, Channel BW/Time/Freq Err, Adj Chan Meas
Time
Mode Channel Analyzer
Dependencies /
Couplings
This key is coupled with the “Frequency Error Limits” key under
the “Meas Setup > Limits >Freq Error Limits” Menu.
Preset Off
State Saved Saved in instrument state.