Chapter 3 163
Mode
Two Port Insertion Loss Measurement
Mode
Marker ->
Accesses the marker function menu keys that change value of center frequency or
reference level to the current value of the selected marker.
Mkr -> RL
Sets the reference level to the amplitude of the specified marker. In delta marker mode,
the reference level is set to the amplitude difference between the markers. (For example, if
the difference is –15 dB, the reference level will bet set to –15 dBm.)
Meas
In the Stimulus Response mode, Meas accesses a menu of keys associated with the
Stimulus Response mode. The measurement runs using the current default setup. Or you
can change settings for your unique measurement requirements. Settings can be changed
using the functions available under the
Meas Setup key.
Meas Setup
In the Two Port Insertion Loss measurement, Meas Setup accesses a menu of keys
associated with this measurement.
Mode SR
Key Path Front-panel key
Mode SR
Remote Command :CALCulate:TPILoss:MARKer[1]|2[:SET]:RLEVel
Example CALC:TPIL:MARK2:SET:RLEV
Key Path
Marker ->, Mkr -> RL
Mode: All
Key Path Front-panel key