Chapter 3 171
Mode
Two Port Insertion Loss Measurement
Mode
Level: 1st Segment Right
Sets the power level, for the 1
st
segment to the Right of the Center segment. This key is only available when 1st
has been selected with the Segments key.
Span: 1st Segment Left
Sets the frequency span for the 1
st
segment to the Left of the Center segment. This key is only available when
1st has been selected with the Segments key.
Mode Stimulus / Response
Notes
Key is only displayed when the selected segment is 1
st
, and the
selected segment side is Left.
Preset Not affected by a preset but set per default to 10 dB after a
power cycling
State Saved Saved in instrument state.
Min –10 dB
Max 150 dB
Key Path
Meas Setup, Limits, Limit Masks, Limit Mask 1, 1st Segment: Left
Level
Default Unit dB
Mode Stimulus / Response
Notes
Key is only displayed when the selected segment is 1
st
, and the
selected segment side is Right.
Dependencies /
Couplings
Value is displayed on the key in the currently selected Y Axis
Unit.
Preset Not affected by a preset but set per default to 10 dB after a
power cycling
State Saved Saved in instrument state.
Min –10 dB
Max 150 dB
Key Path
Meas Setup, Limits, Limit Masks, Limit Mask 1, 1st Segment: Right
Level
Default Unit dB