172 Chapter 3
Mode
Two Port Insertion Loss Measurement
Mode
Span: 1st Segment Right
Sets the frequency span for the 1
st
segment to the Right of the Center segment. This key is only available when
1st has been selected with the Segments key.
2nd Segment
Sets the state of the first segment to On or Off. This key is available only when 2
nd
has been selected with the
Segments key.
Mode Stimulus / Response
Notes
Key is only displayed when the selected segment is 1
st
, and the
selected segment side is Left.
Dependencies /
Couplings
If the Center, 1
st
, and 2
nd
segments are On, the left 2
nd
segment
span adjusts on a change in the left 1
st
segment span.
Preset Not affected by a preset but set per default to 2 MHz after a
power cycling
State Saved Saved in instrument state.
Min 0 Hz
Max Frequency Span Setting
Key Path
Meas Setup, Limit Masks, Limit Mask 1, 1st Segment: Left Span
Default Unit Hz
Mode Stimulus / Response
Notes
Key is only displayed when the selected segment is 2
nd
, and the
selected segment side is Right.
Dependencies /
Couplings
If the Center, 1
st
, and 2
nd
segments On, the right 2
nd
segment
span adjusts on a change in the right 1
st
segment span.
Preset Not affected by a preset but set per default to 2 MHz after a
power cycling
State Saved Saved in instrument state.
Min 0 Hz
Max Frequency Span Setting
Key Path
Meas Setup, Limit Masks, Limit Mask 1, 1st Segment: Right Span
Default Unit Hz