Chapter 3 187
Mode
One Port Insertion Loss Measurement
Mode
Cancel Use this key to abort the Calibration process.
Meas
In the Stimulus Response mode, Meas accesses a menu of keys associated with the
Stimulus Response mode. The measurement runs using the current default setup. Or you
can change settings for your unique measurement requirements. Settings can be changed
using the functions available under the
Meas Setup key.
To configure measurement settings, select
One Port Insertion Loss and then press Meas
Setup
.
Use
Control/Sweep to restart the currently selected measurement and change the sweep
points. To select between continuous or single measurements, press the
Cont or Single
front-panel keys. For information about controlling sweep, refer to “Control/Sweep” on
page 98.
Mode Stimulus / Response
State Saved No
Preset The same value as the Start Freq set for the measurement.
Range 10 MHz to 3.0 GHz (Option 503) / to 6.0 GHz (Option 506)
Dependencies /
Couplings
The Calibration Stop Freq must be set higher that the
Calibration Start Freq. If you set the Calibration Stop Freq
lower then the Calibration Start freq, the start frequency will
automatically be set to a value lower that the stop frequency.
Key Path
Freq/Channel, Calibrate
Mode Stimulus / Response
Key Path
Freq/Channel, Calibrate
Mode Stimulus / Response
State Saved Saved in instrument state.
Range Two Port Insertion Loss | One Port Insertion Loss | Return Loss
| Distance to Fault
Key Path
Meas