Chapter 3 189
Mode
One Port Insertion Loss Measurement
Mode
previous trace data. Pressing the Max Hold key for the selected trace sets the trace type to 
Max Hold, causes the trace to be cleared and causes the Max Hold sequence to be restarted.
Repeat  Sets averaging to the repeat function. Repeat averaging updates the trace after the 
selected number of sweeps has been averaged. It is not updated again until the next set of 
data from the Avg Number of sweeps has been collected.
Repeat Max Hold  Sets averaging to repeat with max hold function. Repeat with max hold 
captures and holds the highest magnitude data at the end of each sweep count. Repeat 
with max hold updates the trace after the selected number of sweeps has been made. It is 
not updated again until the next set of data from the Avg Number of sweeps has been 
collected.
Avg Number 
Sets the number of measurement averages used when calculating the measurement result. 
Clear Calibration 
Clear Calibration deletes the current calibration data.
Key Path: Meas setup, Avg Mode, Max Hold
Key Path: Meas setup, Avg Mode, Repeat
Key Path: Meas setup, Avg Mode, Repeat Max Hold
Mode Stimulus / Response
Preset 4
Key Path
Meas Setup, Avg Number 
Mode Stimulus / Response
Key Path
Meas Setup, Clear Calibration