Chapter 3 221
Mode
Distance To Fault
Mode
Distance To Fault
Distance to Fault Description
A signal is transmitted from the RF OUTPUT connector of the spectrum analyzer to the 
cable-under-test. The signals reflected from faults in the cable are received by analyzer.
In performing this measurement, the analyzer uses frequency domain reflectometry. The 
changing interference of the transmitted and reflected signals contains information about 
the distance to one or more faults. This information can be used to find the physical 
distance to the faults. The distance displayed on the analyzer is the physical distance to 
the probable faults, corrected for the cable loss and velocity factor of the cable.
To configure measurement settings, select 
Distance to Fault and then press Meas Setup. 
Pressing 
Control/Sweep enables you to restart a measurement, refer to “Control/Sweep” on 
page 98.
AMPTD/Y Scale Menu
The following AMPTD/Y Scale keys are for setting up a Distance To Fault measurement. 
The measurement must already be running to access these keys. Press 
Meas, Distance To 
Fault
 to turn on the measurement.
Autoscale 
Executes an automatic scaling of the 
Reference Level setting. Autoscale is a GUI only 
function and has no SCPI command.
Mode Stimulus / Response
State Saved Saved in instrument state.
Key Path
MEAS, Return loss 
Mode Stimulus / Response
Key Path Front-panel key
Mode Stimulus/Response
Key Path
AMPTD Y Scale, Autoscale