Chapter 7 359
IEEE and STATus Commands (Remote Commands Only)
Agilent CSA STATus Register Model
IEEE and STATus Commands (Remote
Commands Only)
Questionable Integrity Negative Transition
This command determines which bits in the Questionable Integrity Condition register will
set the corresponding bit in the Questionable Integrity Event register when the condition
register bit has a negative transition (1 to 0). The variable <integer> is the sum of the
decimal values of the bits that you want to enable.
Questionable Integrity Positive Transition
This command determines which bits in the Questionable Integrity Condition register will
set the corresponding bit in the Questionable Integrity Event register when the condition
register bit has a positive transition (0 to 1). The variable <integer> is the sum of the
decimal values of the bits that you want to enable.
Mode All modes
Remote Command :STATus:QUEStionable:INTegrity:NTRansition
<integer>
:STATus:QUEStionable:INTegrity:NTRansition?
Example STAT:QUES:FREQ:NTR 16 ADC Over Range ‘within range’
will be reported to the Integrity Summary of the Status
Questionable register.
Preset 0
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All modes
Remote Command :STATus:QUEStionable:INTegrity:PTRansition
<integer>
:STATus:QUEStionable:INTegrity:PTRansition?
Example STAT:QUES:INT:PTR 16 ADC Over Range ‘became over
range’ will be reported to the Integrity Summary of the Status
Questionable register.
Preset 32767
Min 0
Max 32767