358 Chapter 7
IEEE and STATus Commands (Remote Commands Only)
Agilent CSA STATus Register Model
IEEE and STATus Commands (Remote
Commands Only)
Questionable Integrity Enable
This command determines which bits in the Questionable Integrity Condition Register will
set bits in the Questionable Integrity Event register, which also sets the Integrity
Summary bit (bit 9) in the Questionable Register. The variable <integer> is the sum of the
decimal values of the bits you want to enable.
Questionable Integrity Event Query
This query returns the decimal value of the sum of the bits in the Questionable Integrity
Event register.
NOTE The register requires that the associated PTR or NTR filters be set
before a condition register bit can set a bit in the event register. The
data in this register is latched until it is queried. Once queried, the
register is cleared.
Mode All modes
Remote Command :STATus:QUEStionable:INTegrity:ENABle <integer>
:STATus:QUEStionable:INTegrity:ENABle?
Example STAT:QUES:INT:ENAB 16 ADC Over Range will be reported
to the Integrity Summary of the Status Questionable register.
Preset 32767
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All modes
Remote Command :STATus:QUEStionable:INTegrity[:EVENt]?
Example STAT:QUES:INT?
Preset 0
SCPI Status Bits/OPC
Dependencies
Sequential command