Chapter 7 357
IEEE and STATus Commands (Remote Commands Only)
Agilent CSA STATus Register Model
IEEE and STATus Commands (Remote
Commands Only)
Questionable Frequency Positive Transition
This command determines which bits in the Questionable Frequency Condition register
will set the corresponding bit in the Questionable Frequency Event register when the
condition register bit has a positive transition (0 to 1). The variable <integer> is the sum of
the decimal values of the bits that you want to enable.
Questionable Integrity Register
Questionable Integrity Condition
This query returns the decimal value of the sum of the bits in the Questionable Integrity
Condition register.
NOTE The data in this register is continuously updated and reflects the
current conditions.
Mode All modes
Remote Command :STATus:QUEStionable:FREQuency:PTRansition
<integer>
:STATus:QUEStionable:FREQuency:PTRansition?
Example STAT:QUES:FREQ:PTR 2 Frequency Reference ‘became
unlocked’ will be reported to the Frequency Summary of the
Status Questionable register.
Preset 32767
Min 0
Max 32767
Mode All modes
Remote Command :STATus:QUEStionable:INTegrity:CONDition?
Example STAT:QUES:INT:COND?
Preset 0
SCPI Status Bits/OPC
Dependencies
Sequential command