356 Chapter 7
IEEE and STATus Commands (Remote Commands Only)
Agilent CSA STATus Register Model
IEEE and STATus Commands (Remote
Commands Only)
Questionable Frequency Event Query
This query returns the decimal value of the sum of the bits in the Questionable Frequency
Event register.
NOTE The register requires that the associated PTR or NTR filters be set
before a condition register bit can set a bit in the event register. The
data in this register is latched until it is queried. Once queried, the
register is cleared.
Questionable Frequency Negative Transition
This command determines which bits in the Questionable Frequency Condition register
will set the corresponding bit in the Questionable Frequency Event register when the
condition register bit has a negative transition (1 to 0). The variable <integer> is the sum
of the decimal values of the bits that you want to enable.
Mode All modes
Remote Command :STATus:QUEStionable:FREQuency[:EVENt]?
Example STAT:QUES:FREQ?
Preset 0
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All modes
Remote Command :STATus:QUEStionable:FREQuency:NTRansition
<integer>
:STATus:QUEStionable:FREQuency:NTRansition?
Example STAT:QUES:FREQ:NTR 2 Frequency Reference ‘regained
lock’ will be reported to the Frequency Summary of the Status
Questionable register.
Preset 0
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command