Chapter 7 355
IEEE and STATus Commands (Remote Commands Only)
Agilent CSA STATus Register Model
IEEE and STATus Commands (Remote
Commands Only)
Questionable Frequency Register
Questionable Frequency Condition
This query returns the decimal value of the sum of the bits in the Questionable Frequency
Condition register.
NOTE The data in this register is continuously updated and reflects the
current conditions.
Questionable Frequency Enable
This command determines which bits in the Questionable Frequency Condition Register
will set bits in the Questionable Frequency Event register, which also sets the Frequency
Summary bit (bit 5) in the Questionable Register. The variable <integer> is the sum of the
decimal values of the bits you want to enable.
Mode All modes
Remote Command :STATus:QUEStionable:FREQuency:CONDition?
Example STAT:QUES:FREQ:COND?
Preset 0
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All modes
Remote Command :STATus:QUEStionable:FREQuency:ENABle <integer>
:STATus:QUEStionable:FREQuency:ENABle?
Example STAT:QUES:FREQ:ENAB 2 Frequency Reference Unlocked
will be reported to the Frequency Summary of the Status
Questionable register.
Preset 32767
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command