354 Chapter 7
IEEE and STATus Commands (Remote Commands Only)
Agilent CSA STATus Register Model
IEEE and STATus Commands (Remote
Commands Only)
Questionable Negative Transition
This command determines which bits in the Questionable Condition register will set the
corresponding bit in the Questionable Event register when the condition register bit has a
negative transition (1 to 0). The variable <integer> is the sum of the decimal values of the
bits that you want to enable.
Questionable Positive Transition
This command determines which bits in the Questionable Condition register will set the
corresponding bit in the Questionable Event register when the condition register bit has a
positive transition (0 to 1). The variable <integer> is the sum of the decimal values of the
bits that you want to enable.
Mode All modes
Remote Command :STATus:QUEStionable:NTRansition <integer>
:STATus:QUEStionable:NTRansition?
Example STAT:QUES:NTR 32 Frequency summary ‘questionable
cleared’ will be reported to the Status Byte Register.
Preset 0
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All modes
Remote Command :STATus:QUEStionable:PTRansition <integer>
:STATus:QUEStionable:PTRansition?
Example STAT:QUES::PTR 32 Frequency summary ‘questionable
cleared’ will be reported to the Status Byte Register.
Preset 32767
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command