Chapter 7 353
IEEE and STATus Commands (Remote Commands Only)
Agilent CSA STATus Register Model
IEEE and STATus Commands (Remote
Commands Only)
Questionable Status Summary bit (bit3) in the Status Byte Register. The variable
<integer> is the sum of the decimal values of the bits you want to enable.
NOTE: The preset condition is all bits in this enable register set to 0. To have any
Questionable Events reported to the Status Byte Register, one or more bits need to be set
to 1. The Status Byte Event Register should be queried after each measurement to check
the Questionable Status Summary (bit 3). If it is equal to 1, a condition during the test
may have made the test results invalid. If it is equal to 0, this indicates that no hardware
problem or measurement problem was detected by the analyzer for the condition bit(s)
enabled.
Questionable Event Query
This query returns the decimal value of the sum of the bits in the Questionable Event
register.
NOTE The register requires that the associated PTR or NTR filters be set
before a condition register bit can set a bit in the event register. The
data in this register is latched until it is queried. Once queried, the
register is cleared.
Mode All modes
Remote Command :STATus:QUEStionable:ENABle <integer>
:STATus:QUEStionable:ENABle?
Example STAT:QUES:ENAB 32 Sets the register so that the Frequency
summary will be reported to the Status Byte Register.
Preset 0
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All modes
Remote Command :STATus:QUEStionable[:EVENt]?
Example STAT:QUES?
Preset 0
SCPI Status Bits/OPC
Dependencies
Sequential command