PN 305892901, Rev S
Specications
Compatible Gases
All specications subject to change without notice.
Air, argon, Air, argon, butane, carbon dioxide, ethane, ethene, Freon 12, Freon
22, Freon 32, Freon 114, Freon 116, Freon 125, Freon 134A, Freon 152, helium,
hydrogren, krypton, natural gas, neon, nitrogen, oxygen, propane, propene,
SF6, and xenon. Carbon dioxide requires a custom Sensor assembly. Contact
the factory to conrm compatibility with other gases.
Read the Dryer Warning in the “Special Warnings and Information”
section in the beginning of this manual for more information.
3050-RM Moisture Analyzer Specications
Specication Description
Range Calibrated range of 0–2500 PPMV.
Limit of Detection 0.1 PPMV nominal.
Accuracy ±10 % of reading OR 0.1 PPMV, whichever is greater.
Response Time Typical 63 % of a step change in either direction in less than
5minutes.
Sample Inlet Pressure
Requirement
138–345 kPa (20–50 PSIG)
Exhaust Pressure 0–103 kPa (0–15 PSIG)
Sample Flow Requirement Approximately 150 SCCM in sample saver mode.
Approximately 1 SLPM with bypass open.
Inlet Gas Temperature 0 °C to 100 °C (32 °F to 212 °F)
Optimal results are obtained when inlet gas temperature is
maintained at 60 °C (140 °F).
Outputs 4-line x 20-character LCD display.
One (1) 4–20 mA DC analog output.
RS-232/RS-485 serial port.
Specications | 5-1