Special Addresses
Addresses $F0 through $FF are reserved for special functions with address
$F0 used as a broadcast address.
Address $F0
Node address $F0 is used as a single point address (i.e., it is intended for use in
congurations with a single slave). The slave will respond to all messages with
node address $F0. You must be careful to prevent multiple slaves from receiv-
ing commands with the broadcast address as this will cause a bus contention.
Dened Commands (Master to Slave)
The commands that can be issued from the master are listed in Figure 2-19.
Command Byte Function
A Echo
B Bad Command
C Acknowledge
F Read Data
G Start Verify
H Write Data
L Write Cell EEPROM
Q Quit Verify
R Analyzer Reset
T Toggle Test Mode
Figure 2-19.
Dened Commands table.