Sample Gas
This list allows you to specify the Sample gas you want to analyze.
Select Gas
Press Enter to move to the listing of Sample gases. Scroll up/down the list
to select the correct Sample gas.
Custom Gas
If you select the Custom Gas selection and press Enter, you will be
prompted to enter a Gas K Factor. The Gas K Factor is a parameter used
by the Flow Meter in the 3050-RM to adjust for the molar heat capacities of
dierent gases. Nitrogen is used as the Reference gas for the K factors, and
its value is dened as 1.000. To estimate the K factor for another gas, take
the ratio of the molar heat capacity (constant pressure) of nitrogen to that
of the sample gas. For example, the K factor of argon would be estimated
as follows:
Molar heat capacity of nitrogen 29.12 JK
-1
mole
-1
Molar heat capacity of argon 20.79 JK
-1
mole
-1
K factor 29.12 JK
-1
mole
-1
/20.79 JK
-1
mole
-1
K factor 1.401
Config
Sample Gas
Select Gas
Air
Argon
Butane
CarbonDioxide
CustomGas
Ethane
Ethene
Freon12
Freon22
Freon32
Freon114
Freon116
Freon125
Freon134A
Freon152
Helium
Hydrogen
Krypton
NaturalGas
Neon
Nitrogen
Oxygen
Propane
Propene
SF6
Xenon
Gas K Factor
Sensor Saver
Enable
Disable
Gas Saver
Enable
Disable
Figure 3-6.4.
Sample Gas sub-menu.
3-8 | 3050-RM Moisture Analyzer