PN 305892901, Rev S
Analyzer Installation
Consult plant safety personnel for appropriate exhaust venting guide-
lines for specic Sample gas type.
User-Supplied Items Interconnect Wiring
The user must supply all wiring for serial communication, analog output, and
alarm relays. Recommended wire is 22 gauge, twisted/shielded pair.
External Sample Tubing
 Sample input tubing length should be as short as possible.
- For optimum performance, sample tubing should be thermally con-
trolled to maintain a constant Sample Inlet temperature. Optimal
results are obtained with the sample line maintained at 60 °C.
- Recommended Sample gas input tubing is 1/8-inch OD, electropol-
ished 316L VAR stainless steel (AMETEK Part No. 257707000) or equiva-
lent.
Exhaust pressure must be 0–15 PSIG.
2-6 | 3050-RM Moisture Analyzer