PN 305892901, Rev S
Installing the Dryer
Prior to operating the 3050-RM Analyzer, the Dryers must be installed. One
(1) Dryer has been supplied with this analyzer. Make sure that you have the
new Dryer and the new VCR gaskets available before beginning installation.
The amount of time that the Dryer ttings are left open should be kept
to less than one (1) minute to avoid damage to the device.
To install the Dryer (see Figures 2-4.1 and 2-4.2):
1. On the right side of the rear of the analyzer is the Dryer access compartment
(Figure 2-4.1).
2. Loosen the four (4) captured screws on the door of the compartment to
remove the door.
3. If you have a single Dryer analyzer, you must install the Dryer in the bot-
tom position.
An optional second Dryer or contaminant trap is available. If you have
the two-dryer or dryer/trap version of the 3050-RM, follow the instruc-
tions for the one-dryer operation and then repeat for the second dryer
or trap.
Figure 2-4.1.
Dryer location for installation.
Installation and Start-Up | 2-9