PN 305892901, Rev S
Specication Description
Alarms/Alerts Three (3) independent contact closures of 30 VAC or 60 VDC
maximum, 10 VA or 1 A maximum resistive, for System alarm,
moisture Concentration Alarm, and Data Valid. All are fail-safe by
default.
Alarms are available on RS-485 interface.
Environmental Conditions 5 °C to 50 °C (41 °F to 122 °F)
[Optimal performance in low-level applications is obtained when
ambient temperature is maintained at a constant temperature
(±2 °C)]
90 % RH maximum, noncondensing
IEC Pollution Degree 2
Maximum altitude 2,000 meters (6560 feet)
IEC Installation Category II
Indoor use only, IEC 529; IP40
Utility Requirements 100127 VAC 4763 Hz, 185 W max
OR
220240 VAC, 4763 Hz, 185 W max
Mounting 19-inch rack
Dimensions Width: 48 cm (19 inches)
Height: 13 cm (5.25 inches)
Depth: 42 cm (16.5 inches)
Net Weight 10.9 kg (24 lb)
3050-RM Moisture Analyzer Approvals and Certications
Design Classications MET Certied to: UL/CSA
General Safety Requirements: UL61010, CAN/
CSA C22.2 No. 1010.1
Complies with: UL1604 and CAN/
CSA C22.2 No. 213
Class 1, Division 2, Groups A,B,C,D T4
EMC Directive: EN61326-1
Industrial Low Voltage Directive: EN61010-1
5-2 | 3050-RM Moisture Analyzer