PN 305892901, Rev S
GROUND STUDS
DRYER COMPARTMENT
COVER
RS-232
Instrument Air
80 - 100 PSIG
Exhaust
Out
Atmospheric
Pressure
Sample In
20-50 PSIG
Figure 2-8.
Attach ground wire to
threaded chassis Ground Stud.
6. Feed the strain relief cap with compression bushing onto the new cord and
tighten securely as per original. If the new cord is a smaller diameter than
the original, consider using the alternate bushing that is included with the
European cord. Ensure that the power cord does not slip when pulled or
pushed from the outside.
7. Reconnect 1 (L) and 2 (N) wires (Figure 2-7) by screwing in the two (2) plug
screws. Make the ground connection to the threaded chassis Ground Stud
(Figure 2-8). Do not use the ground position on the power connector. Plug
in the power connector.
8. Connect a proper, locally recognized plug or terminator to the end of your
cord.
9. Replace the Rear Panel.
10. Reconnect power.
2-14 | 3050-RM Moisture Analyzer