EasyManua.ls Logo

Ametek 3050-RM - Page 76

Ametek 3050-RM
111 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PN 305892901, Rev S
Set Dryer Production Code button
Click to enter the Dryer Production Code for the dryer that is installed in
the analyzer. Click OK to accept.
You must enter a new dryer code each time you replace the dryer.
Set Moisture Generator Production Code
Click to enter the Moisture Generator Production Code.
Click OK to accept.
Figure 3-11.2.
Dryer Production Code entry
dialog box.
Figure 3-11.3.
Moisture Generator Production
Code entry dialog box.
Verication Duration
Enter the Verication duration in minutes. The system defaults to the mini-
mum time required.
Scheduled Verication
Schedule routine Verications by clicking the Daily, Weekly, or Monthly
button.
Never
No verify will be done. Scheduling slots at the bottom of the box are
not available.
3-24 | 3050-RM Moisture Analyzer