Chapter 7 Remote Commands
7-6
:DISPlay:ACTive <unit>,<slot>[,<tab>]
Parameter <unit>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Mainframe No.1 to 4
When installing the MT1810A, the mainframe No. is specified.
When using the MP1800A, No.1 is fixed.
0 When specifying a USB module
(MP1825B/MP1821A/MP1822A/MP1861A/MP1862A)
<slot>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 6 Slot No.1 to 6
When using the MP1800A, the slot No. is 1 to 6.
When using the MT1810A, the slot No. is 1 to 4.
1 to 14 USB No.1 to 14
When using a USB module (<unit>=0), set the USB No.1
to 14.
[<tab>]=<DECIMAL NUMERIC PROGRAM DATA>
1 to X Tab ID No.1 to X
When [,<tab>] is omitted, 1 is specified.
Tab ID is set to No.1 at the left side dialog, and then the tab ID No. is set to
2, 3, 4...toward the right side.
The maximum number (X) of the tab ID varies depending on the module
options.
Function Displays the specified module screen to the front.
Note:
This command is not available if the measurement result display
processing is set to Off.
To use the measurement result display processing, set it to On
with the :SYSTem:DISPlay:RESult ON command.
Example To display the Interface tab of MP1862A:
(When MP1862A is connected to USB11)
> :DISP:ACT 0,11,1
Compatibility Incompatible with existing models.