7.3 64G DEMUX Commands
7-27
7
Remote Commands
:DEMux:CLOCk:RELative? [<interface>]
Response <boolean>=<NR1 NUMERIC RESPONSE DATA>
0 Reference OFF
1 Reference ON
Function Queries clock input phase variation reference setting.
Example To query Data Interface clock input phase variation reference:
> :DEMux:CLOCk:RELative?
< 1
Compatibility Compatible with MU182040A and MU182041A.
:DEMux:CLOCk:RDELay <numeric>,<unit>
Parameter <numeric>=<DECIMAL NUMERIC PROGRAM DATA>
(mUI units) –2000 to 2000 –2000 to 2000 mUI /4 mUI Step
(ps units) Converted from the calculated frequency value, based on the
setting resolution of mUI.
<unit>=< CHARACTER PROGRAM DATA>
UI mUI units
PS ps units (The ps unit is selected when <unit> is omitted.)
Function Sets clock input phase variation reference value and units.
Setting is performed as an increment from the reference value.
The specified value may not be exactly set as is when the ps unit is set, due
to the limitation of the setting resolution.
The value is adjusted to the most approximate value and set in this event.
Example To set Data Interface clock phase variation reference to –1000 mUI:
> :DEMux:CLOCk:RDELay -1000,UI
Compatibility Compatible with MU182040A and MU182041A.
:DEMux:CLOCk:RDELay? [<unit>]
Parameter <unit>=< CHARACTER PROGRAM DATA>
UI mUI units
PS ps units (The ps unit is selected when <unit> is omitted.)
Response <numeric>=<NR1 NUMERIC RESPONSE DATA>
Function Queries clock input phase variation reference value and units.
Example To query Data Interface clock phase variation reference UI units:
> :DEMux:CLOCk:RDELay? UI
< -1000
Compatibility Compatible with MU182040A and MU182041A.